Torque Standard Machines

To calibrate a transducer you need a standard machine to load it. Depending on requirements and measuring range we recommend the following standard machines:
  • Lever and deadweight
    A variable mass stack loads the transducer via a lever of a constant length.
  • Jockey weight
    A constant mass movable on the lever loads the transducer via the lever.
  • Lever and force reference
    A force, controlled by a force measuring system, loads the transducer via a lever of a constant length.